DHR/AR Rheometer Error Messages

Refer to the table below for the list of DHR/AR Series Rheometer error messages and possible solutions.

 

Message Code Message Text Explanation of Message Solutions to Error
30001 Unrecognized command, invalid number of parameters or invalid parameter value (COMERR) Command sent - <Command> The instrument did not recognize the command sent to it, or the values passed were out of range.  
30003 Instrument overspeed (SPDERR) Shown when the rheometer bearing rotation exceeds the specified maximum speed.  
30005 Fault or excessive force detected when attempting to zero normal force reading (NFERR) This message is displayed if an excessive normal force is detected when attempting to zero the normal force reading. This could be caused by either an excessive force being applied by air bearing/sample/mechanical interference with ETC or a gauge fault.

Make sure there is nothing touching the ETC.

Contact TA Instruments' Service Department.

30006 Stop button pressed while front panel locked (OPERR) This message is displayed when the Stop button is pressed while the TRIOS software has the keypad locked. This usually indicates that a test run or calibration is in progress and the operator aborted it using the rheometer stop button.  
30007 Bearing pressure low (BPLERR) This is displayed if the air supply is inadvertently switched off while the rheometer is on or if the supply pressure has dropped below the minimum operating pressure. Ensure that regulator is indicating the correct pressure.
30008 Auto gap set error (AGERR) Error is displayed during zero gap routine if reference position from 2nd or 3rd try is significantly different from 1st try. Clean plates and check for damage. If error occurs during deceleration mode, it could be due to a failing air bearing.
30009 Transducer initialisation failure (TRANERR) The bearing is unable to rotate during instrument start up. Remove bearing cap. Make sure air is supplied.
30010 Peltier overheat (PELTERR)   Ensure that cooling fluid is being circulated.
30011 PT100 error (PTAERR)   Contact TA Instruments' Service Department.
30012 PT100 error (PTBERR)   Contact TA Instruments' Service Department.
30013 PT100 error (PTCERR)   Contact TA Instruments' Service Department.
30014 PT100 error (PTDERR)   Contact TA Instruments' Service Department.
30015 Cable fault (CABERR)   Contact TA Instruments' Service Department.
30016 Safety interlock open (INTERR)   Contact TA Instruments' Service Department.
30017 Overheat error (OVHTERR)   Contact TA Instruments' Service Department.
30018 General temperature system error (GTERR)   Contact TA Instruments' Service Department.
30019 PT100 error (PT100ERR)   Contact TA Instruments' Service Department.
30021 Initializing processor board (EPINIT) Information NA
30022 Error on processor board (EPERR)   Contact TA Instruments' Service Department.
30023 Initializing power board (PPINIT) Information NA
30024 Error on power board (PPERR)   Contact TA Instruments' Service Department.
30025 Initializing instrument section (IPINIT) Information NA
30026 Error in instrument section (IPERR)   Contact TA Instruments' Service Department.
30027 Power cable not correctly connected (PWRCAB) This indicates that the power cable may not be firmly plugged in. Plug in the power cable.
30028 Signal cable not correctly connected (SIGCAB) This indicates that the power cable may not be firmly plugged in. Plug in the power cable.
30029 Power board heatsink overheat (OTERR)   Turn off instrument and ensure that fans are clear and that air is free to circulate.
30030 Temperature system sensor error (SENSOR)   Contact TA Instruments' Service Department.
30031 Temperature system element error (ELEMENT)   Contact TA Instruments' Service Department.
30032 Temperature system environment error (e.g. no peltier water) (ENVIRON)   Ensure that all accessories are correctly connected and configured.
30033 Pre hardware initialisation (only at system startup) (PREINIT)   Contact TA Instruments' Service Department.
30034 Fault or extreme temperature on thermocouple cold junction sensor (CJCERR)   Contact TA Instruments' Service Department.
30035 Fault or extreme temperature on normal force temperature sensors (NFPTERR) This message is displayed if there are faulty normal force temp sensors or if, for example, the instrument is frozen due to excessive liquid nitrogen. Let instrument equilibrate to room temperature.
30036 Controlled rate instability (CRERR)   Check that instrument and geometry inertia values are correct.
30037 The encoder linearisation must be done before the instrument can be used. Information Perform linearization.
30038 Power fail (PFERR) Displayed if the instrument is turned off while the software is communicating with it.  
30039 Internal system fault (SYSTEMERR)   Contact TA Instruments' Service Department.
30040 Magnetic bearing unstable/position lost (MAGBERR) Often the result of a sample volume change when axial force control is not being used. Change procedure, including axial force control.
30041 Startup self test error (POSTERR)   Contact TA Instruments' Service Department.
30042 Maximum motor temperature exceeded (MOTORTP) The maximum motor temperature has been exceeded. Reduce the torque being used in the procedure. For DHR, supply motor cooling.
31000 Invalid IP address <IP address> An invalid IP address has been entered. If a command line has been used explicitly, check the IP address.
31001 Failed to connect to instrument at <IP address> <error message>   Check that the instrument is switched on.
31002 WaitTimeout <Command> Failed to send the command to the instrument because another command was already waiting. Contact TA Instruments' Service Department.
31003 The instrument failed to respond before a timeout occurred - Command <Command>  
  • Check that the instrument is still running.
  • Ensure that the network is functioning properly.
  • Restart TRIOS.
31004 Failed to send the command to the instrument - Command <Command> <Error message>  
  • Check that the instrument is still running.
  • Ensure that the network is functioning properly.
  • Restart TRIOS.
31005 Failed to receive the acknowledgement – Command <Command>  
  • Check that the instrument is still running.
  • Ensure that the network is functioning properly.
  • Restart TRIOS.
31006 The comms reply was invalid - Command <Command> Reply : <Reply>   Contact TA Instruments' Service Department.
31007 The reply was the same as the command <Command>   Contact TA Instruments' Service Department.
31008 Experiment aborted by user Information NA
31011 Failed to read reply from instrument - Command <Command> <error message> Failed to read a reply from the instrument because of a socket failure. If no data in the Control panel is being updated, then restart TRIOS.
31012 The socket is not connected - Command <Command> The socket is not connected when trying to send a command to the instrument. If no data in the Control panel is being updated, then restart TRIOS.
31013 The socket disconnected while reading - Command <Command> The socket disconnected while reading from the instrument. If no data in the Control panel is being updated, then restart TRIOS.
32000 The calculated friction value <value> is not within the required range. This may be because the spindle cannot rotate freely.   Check that the bearing is free to rotate.
32001 The friction cannot be calculated with a zero or negative velocity. Unexpected value following calibration. Contact TA Instruments' Service Department.
32004 Instrument inertia calibrated : New value <inertia value> Information NA
32005 Geometry inertia calibrated : New value <inertia value>, Geometry : <geometry name> Information NA
32006 Compliance calibrated : New value <value>, Geometry : <geometry name> Information NA
32007 Gap temperature compensation calibrated : New value <value>, Geometry : <geometry name> Information NA
32008 Friction calibrated : New value <value>, Geometry : <geometry name> Information NA
32009 UV irradiation calibrated : New value <value> Information NA
32010 UHP Zero Heat Flow calibrated : Gradient Span <span value>, offset <offset value> Information NA
32011 EHP Zero Heat Flow calibrated : Gradient Span <span value>, offset <offset value> Information NA
32012 Encoder linearization done Information NA
32013 Auto asphalt temperature calibrated : Span <span value>, offset <offset value> Information NA
32014 Rotational mapping done - <geometry name> Information NA
32015 Bearing test run Information NA
32016 Oscillatory mapping done : <mapping description> Information NA
32018 Instrument name changed to <new name> Information NA
32019 Instrument location changed to <new location> Information NA
32020 Smart swap geometry serial number programmed (<serial number>) Information NA
32021 The instrument could not read the smart swap geometry serial number <error>.   Contact TA Instruments' Service Department.
32022 The spindle cannot be turned so the smart swap serial number cannot be programmed. The geometry Smart Swap needs to be able to turn the spindle to allow the serial number to be read. Raise the instrument head and check to make sure that the spindle can rotate freely.
32023 Reading the smart swap serial number. Please wait... Information NA
32024 Smart swap read disabled Information NA
32025 No geometry fitted Tried to read geometry Smart Swap but no geometry was fitted. Fit the geometry.
32026 Instrument starting up The geometry Smart Swap cannot be read until the instrument has finished starting up.  
32030 The calculated gap temperature compensation value <value> is not within the required range. During gap temperature compensation calibration, the calculated value is out of range. Contact TA Instruments' Service Department.
32031 Inertia calibration failed - zero values encountered. The frequency or displacement value read back from the instrument is zero. Contact TA Instruments' Service Department.
32032 The calibration has been abandoned because the gradient offset has become too large. During UHP calibration. Contact TA Instruments' Service Department.
32033 The calculated offset value <value> is not within the required range. During auto asphalt temperature calibration, the calculated offset value is out of range. Contact TA Instruments' Service Department.
32034 The calculated span value <value> is not within the required range. During auto asphalt temperature calibration, the calculated span value is out of range. Contact TA Instruments' Service Department.
32035 Encoder linearization - external program failed to connect to the instrument.   Contact TA Instruments' Service Department.
32036 Encoder linearization - bad command line   Contact TA Instruments' Service Department.
32037 Encoder linearization - socket error   Contact TA Instruments' Service Department.
32038 Encoder linearization - external program missing <file name>   Contact TA Instruments' Service Department.
32039 Encoder linearization - failed to delete the response file - <file name>  
  • Check that the file is not read-only.
  • Contact TA Instruments' Service Department.
32040 Encoder linearization - failed to create the directory for the response file <directory name> - <error message>  
  • Check that the directory has create rights.
  • Contact TA Instruments' Service Department.
32041 Encoder linearization - failed to run the external program - <file name>   Contact TA Instruments' Service Department.
32042

Encoder linearization - unknown error <error code>
<error text>

  Contact TA Instruments' Service Department.
32043 Encoder linearization - failed to read the response file <file name> - <error>  
  • Check that the file is not read-only.
  • Contact TA Instruments' Service Department.
32050 Failed to create the voltmeter object Auto asphalt temperature calibration. Contact TA Instruments' Service Department.
32051 No digital voltmeter found. Auto asphalt temperature calibration. Connect the digital voltmeter to the PC and switch it on.
32052 Digital voltmeter found but device has not been put into ohms mode. Auto asphalt temperature calibration. Put the device into ohms mode.
32053 The DigitalVoltmeter.dll has not been installed Auto asphalt temperature calibration. Contact TA Instruments' Service Department.
32060 The wizard cannot load the sample as the gap has not been zeroed. Submersion clamps sample loading. Zero the gap from the Control panel.
32061 The wizard cannot position the head as the gap stored after mounting the sample is invalid. Submersion clamps sample loading. Try loading again, following the instructions closely.
32070 Failed to write the addresses to the instrument.   Contact TA Instruments' Service Department.
32071 Failed to reconnect to the instrument after changing the network addresses.   Contact TA Instruments' Service Department.
32072 The instrument's network addresses were changed. Information NA
32073 Invalid password   The password should be less than 65 characters.
32074 Failed to set the password in the instrument.   Contact TA Instruments' Service Department.
33000 <error message> A communications error occurred when checking the geometry at the start of the test. Contact TA Instruments.
33001 The selected geometry is the submersion clamps solid sample which must be used with the Peltier concentric cylinders. The experiment cannot continue Shown at the start of an experiment. Change the selected geometry.
33002 Warning: The instrument is set to use the ETC in plate mode but the selected geometry is not an ETC plate Shown at the start of an experiment – the option is given to continue. Continue with the experiment or change the selected geometry.
33003 Warning: The instrument is set to use a solid sample but the selected geometry is not solid sample clamps Shown at the start of an experiment – the option is given to continue. Continue or correct the compatabitity identified.
33005 Failed to read the UV type <error message>   Contact TA Instruments' Service Department.
33006 Failed to read smart swap serial number <error message> Smart Swap could not be read.

Check integrity of Smart Swap sleeve. If damaged, it will need to be changed. Contact TA Instruments' Service Department.

33007 The experiment cannot be run because the loaded geometry does not match the one attached to the instrument   Switch off Smart Swap from the geometry menu if you want to run using the current setup
33008 <error message> <error code> One of the geometry dimensions is invalid. Correct the dimension.
33009 The geometry is not for this instrument type.   Select another geometry.
33010 There is no geometry selected   Select a geometry.
33011 The <step name> step cannot be run using the current geometrySelect another geometry (at present the only restriction is an extensional fixture must be used with an extensional step)   Select another geometry (at present, the only restriction is an extensional fixture must be used with an extensional step).
33100 Firmware updated to version <version> Information NA
33101 The instrument's network addresses were changed. Information NA
33102 Instrument password changed. Information NA
33103 Failed to reconnect to the instrument after upgrading the firmware.   Close TRIOS and then re-open the application.
33104 The firmware file was not found - <file name>   Contact TA Instruments' Service Department.
33200 A zero value is not allowed for the start or end value for a log ramp OR the start and end values for a log ramp must have the same sign. This will occur if a log ramp uses an inherited start value and either that inherited value is 0 or the inherited value does not match the sign of the end value.  
33202 Transient data file <filename> failed - <error message> A failure occurred during a transient oscillation test ( time sweep or amplitude sweep) – the error message will provide additional information. Contact TA Instruments' Service Department.
33203 Transient data collection failed - <error message> A failure occurred during a transient oscillation test ( time sweep or amplitude sweep) – the error message will provide additional information. Contact TA Instruments' Service Department.